Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0348-2027

restor3d Veritas Reverse Total Shoulder Replacement Recall

Veritas Reverse Total Shoulder Replacement, Model/Catalog Number: 3500-140C3, 3500-140U3, 3500-140C1, 3500-140U1, size 1 -2, 40mm Polymer Liner

FDA recall Class II

Ongoing

restor3d recalled the listed Veritas shoulder components because a particular stem-size and glenosphere pairing can create excessive mechanical loading. FDA’s notice reports a potential loss of retention strength in certain polyethylene components.

Recall at a glance

Recall number
Z-0348-2027
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Oct 1, 2026
FDA report date
Nov 11, 2026
Quantity in commerce
240 units
Distribution
See key details
FDA cause
Device Design

What is being recalled?

Veritas Reverse Total Shoulder Replacement, Model/Catalog Number: 3500-140C3, 3500-140U3, 3500-140C1, 3500-140U1, size 1 -2, 40mm Polymer Liner

Affected model & UDI checker

Search the exact identifiers published for Z-0348-2027.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
3500-140C3Veritas Reverse Total Shoulder ReplacementAll lot codes listed for this model00840097513492
3500-140U3Veritas Reverse Total Shoulder ReplacementAll lot codes listed for this model00840097513461
3500-140C1Veritas Reverse Total Shoulder ReplacementAll lot codes listed for this model00840097513430
3500-140U1Veritas Reverse Total Shoulder ReplacementAll lot codes listed for this model00840097513409
FDA code information

All Lot codes/ UDIs:Product Code: 3500-140C3 UDI/GTIN:00840097513492; Product Code: 3500-140U3 UDI/GTIN:00840097513461; Product Code: 3500-140C1 UDI/GTIN:00840097513430; Product Code: 3500-140U1 UDI/GTIN:00840097513409

View the original FDA recall notice

Why was this product recalled?

(Retroactively reported) Pairing of shoulder implant stem sizes 1 and 2 with the largest glenosphere diameter (40 mm) creates an excessive mechanical loading configuration in which certain polyethylene components lack adequate retention strength, potentially resulting in dislocation, reduced range of motion, or revision surgery.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did restor3d Inc. take?

FDA reports that restor3d contacted affected consignees by telephone and emailed an urgent medical-device recall notice on October 7, 2026. The firm’s listed instructions were to:

  1. Not implant or otherwise use the identified affected units.
  2. Return the Field Action Response Form to restor3d within 10 business days.
  3. Promptly report intraoperative issues, suspected device complications and the other clinical concerns specified in the notice to restor3d.
Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. Urgent recall notice emailed to consignees

  3. FDA notice posted

  4. FDA classification recorded

  5. FDA enforcement report date

Related FDA submissions

Product / regulatory recordCompanyPathwayDecision dateSource
Restor3D510(k) clearanceMay 19, 2025FDA

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology