restor3d Veritas Reverse Total Shoulder Replacement Recall
Veritas Reverse Total Shoulder Replacement, Model/Catalog Number: 3500-140C3, 3500-140U3, 3500-140C1, 3500-140U1, size 1 -2, 40mm Polymer Liner
FDA recall Class II
Ongoing
restor3d recalled the listed Veritas shoulder components because a particular stem-size and glenosphere pairing can create excessive mechanical loading. FDA’s notice reports a potential loss of retention strength in certain polyethylene components.
Recall at a glance
- Recall number
- Z-0348-2027
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Oct 1, 2026
- FDA report date
- Nov 11, 2026
- Quantity in commerce
- 240 units
- Distribution
- See key details
- FDA cause
- Device Design
What is being recalled?
Veritas Reverse Total Shoulder Replacement, Model/Catalog Number: 3500-140C3, 3500-140U3, 3500-140C1, 3500-140U1, size 1 -2, 40mm Polymer Liner
Affected model & UDI checker
Search the exact identifiers published for Z-0348-2027.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 3500-140C3 | Veritas Reverse Total Shoulder Replacement | All lot codes listed for this model | 00840097513492 |
| 3500-140U3 | Veritas Reverse Total Shoulder Replacement | All lot codes listed for this model | 00840097513461 |
| 3500-140C1 | Veritas Reverse Total Shoulder Replacement | All lot codes listed for this model | 00840097513430 |
| 3500-140U1 | Veritas Reverse Total Shoulder Replacement | All lot codes listed for this model | 00840097513409 |
FDA code information
All Lot codes/ UDIs:Product Code: 3500-140C3 UDI/GTIN:00840097513492; Product Code: 3500-140U3 UDI/GTIN:00840097513461; Product Code: 3500-140C1 UDI/GTIN:00840097513430; Product Code: 3500-140U1 UDI/GTIN:00840097513409
Why was this product recalled?
(Retroactively reported) Pairing of shoulder implant stem sizes 1 and 2 with the largest glenosphere diameter (40 mm) creates an excessive mechanical loading configuration in which certain polyethylene components lack adequate retention strength, potentially resulting in dislocation, reduced range of motion, or revision surgery.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did restor3d Inc. take?
FDA reports that restor3d contacted affected consignees by telephone and emailed an urgent medical-device recall notice on October 7, 2026. The firm’s listed instructions were to:
- Not implant or otherwise use the identified affected units.
- Return the Field Action Response Form to restor3d within 10 business days.
- Promptly report intraoperative issues, suspected device complications and the other clinical concerns specified in the notice to restor3d.
Recall timeline
Recall initiated by firm
Urgent recall notice emailed to consignees
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-0348-2027
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 19, 2026
- openFDA device recall record · Source release Nov 14, 2026 · Retrieved Nov 19, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology